Ap. Argekar et al., SIMULTANEOUS DETERMINATION OF MEBENDAZOLE AND PYRANTEL PAMOATE FROM TABLETS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY REVERSE-PHASE (RP-HPLC), Talanta, 44(11), 1997, pp. 1959-1965
A new, simple, precise, rapid and stability indicating RP-HPLC method
has been developed for the simultaneous determination of mebendazole a
nd pyrantel pamoate from tablets. Chromatography was carried out on a
Shodex C8 column using a mixture of 0.05 M monobasic sodium phosphate:
acetonitrile: triethylamine (60:40:1.5, v/v), adjusted to pH 6.8 with
phosphoric acid. Detection was carried out at 290 nm using a UV detec
tor. Retention times for pamoic acid, pyrantel base and mebendazole we
re 3.61, 6.41 and 12.81 min, respectively. Linearity was obtained in t
he concentration range of 20-160 mu g and 30-240 mu g for pyrantel pam
oate and mebendazole, respectively. The method after its application f
or the assay of pyrantel pamoate and mebendazole from tablets, was sta
tistically evaluated for its accuracy and precision. (C) 1997 Elsevier
Science B.V.