A STEREOTAXIC TARGETING DEVICE IN VACUUM-CORE BREAST BIOPSY - A NEW METHOD IN PERCUTANEOUS DIAGNOSTIC-BIOPSY DEMONSTRATED IN 120 CASES

Citation
C. Kuchler et al., A STEREOTAXIC TARGETING DEVICE IN VACUUM-CORE BREAST BIOPSY - A NEW METHOD IN PERCUTANEOUS DIAGNOSTIC-BIOPSY DEMONSTRATED IN 120 CASES, Radiologe, 37(8), 1997, pp. 621-628
Citations number
22
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
0033832X
Volume
37
Issue
8
Year of publication
1997
Pages
621 - 628
Database
ISI
SICI code
0033-832X(1997)37:8<621:ASTDIV>2.0.ZU;2-Q
Abstract
Purpose: To test the capabilities of vacuum core biopsy (VCB) in the d iagnosis of mammographically indeterminate lesions. Materials and Meth ods: 120 patients (131 lesions) were examined using VCB with 14G or 11 G vacuum core. Results: VCB was mostly performed because of indetermin ate microcalcifications (67 cases) or soft tissue densities/architectu ral distortion (64 cases). 112 benign changes, 14 DCIS and 5 invasive carcinomas were found. Excellent accuracy was achieved (presently 100 %), since complete excision of small lesions/areas (less than or equal to 1 cm) or partial excision of larger lesions was possible. No relev ant hematomas or infections occurred. Patients tolerated the painless procedure very well. Discussion: This report confirms our previous exp eriences. This method promises to replace diagnostic open biopsy of in determinate or suspicious nonpalpable lesions.