H. Shi et al., SULFUR-SELECTIVE CHEMILUMINESCENCE DETECTION WITH PACKED-COLUMN SUPERCRITICAL-FLUID CHROMATOGRAPHY, Journal of chromatography, 779(1-2), 1997, pp. 307-313
Citations number
13
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
A new generation sulfur chemiluminescence detection (SCLD) system was
interfaced and tested for supercritical fluid chromatography (SFC) wit
h packed columns using 100% SF-CO2 and methanol modified CO2 as the mo
bile phase. The detection chemistry for the SCLD is based on ozone-ind
uced chemiluminescence following a two-step combustion process of cons
ecutive oxidation and reduction of sulfur-containing compounds. A seve
n-day evaluation study showed excellent sensitivity, selectivity and l
inearity, as well as day-to-day repeatability. The minimum detectable
quantity was determined to be 3 pg sulfur (0.2 pg S/s) at the detector
. Equimolar response of SCLD to sulfur compounds with different bondin
g environments was also observed. Unique applications and capabilities
of the SFC-SCLD system for sulfur speciation and detection are presen
ted for a petroleum product, thermally labile pesticides and herbicide
s which are difficult or impossible to analyze by GC techniques, (C) 1
997 Elsevier Science B.V.