Th. Eichhold et al., HIGHLY SENSITIVE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC TANDEM MASS-SPECTROMETRIC METHOD FOR THE ANALYSIS OF DEXTROMETHORPHAN IN HUMAN PLASMA, Journal of chromatography B. Biomedical sciences and applications, 698(1-2), 1997, pp. 147-154
Citations number
17
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
A stable-isotope-dilution HPLC-tandem mass spectrometry-based method w
as developed for the determination of dextromethorphan in human plasma
. Plasma samples were prepared for analysis by solid-phase extraction
on octadecylsilane extraction cartridges. Dextromethorphan and the deu
terium-labeled dextromethorphan internal standard were chromatographed
on a short reversed-phase column and detected by a selected-reaction-
monitoring scheme. Linear standard curves were obtained over three ord
ers of magnitude and the limit of quantitation for dextromethorphan wa
s 50 pg/ml, using a 1-ml plasma sample. The combination of HPLC and el
ectrospray tandem mass spectrometry resulted in a rapid, selective and
sensitive method for the analysis of dextromethorphan in plasma. The
method was applied for the evaluation of the pharmacokinetic profile o
f dextromethorphan in human volunteers following peroral administratio
n. (C) 1997 Elsevier Science B.V.