Pa. Lonnqvist et B. Lofqvist, DESIGN FLAW CAN CONVERT COMMERCIALLY AVAILABLE CONTINUOUS SYRINGE PUMPS TO INTERMITTENT BOLUS INJECTORS, Intensive care medicine, 23(9), 1997, pp. 998-1001
Objective:To investigate if unexpected behaviour of neonatal and paedi
atric patients connected to syringe pumps could be explained by transi
ent elevation of these devices. Design: Five different commercially av
ailable syringe pumps were set at an infusion rate of 1 ml/h and then
subjected to a vertical displacement manoeuvre (height 1 m). The actua
l delivered infusion volumes in association with the displacement mano
euvre were measured by a high precision weight scale connected to a co
mputer. Setting: A medical technology laboratory in a university hospi
tal. Measurements and results: Elevation of the devices resulted in a
rapid bolus injection of 0.19-2.28 ml. Returning the devices to their
original positions resulted in an aspiration into the system of 0.06-0
.34 ml. The times both for bolus injection and for aspiration into the
system were less than 1 min in all cases. The up-down manoeuvre was f
ollowed by a period with zero infusion ranging from 8 to 105 min. Conc
lusions: Design flaws in the construction of syringe pumps can expose
patients to substantial danger following vertical displacement if pote
nt drugs are being infused. If potent drugs are infused, care should b
e taken not to change the vertical position of the syringe pump even f
or short periods of time. Before buying new equipment, the authors rec
ommend that the delivery characteristics of these devices should not o
nly be tested during ordinary bench testing but should also include th
e reaction to a vertical displacement manoeuvre.