Pcm. Bartels et al., EVALUATION OF SYSMEX SF-3000 PERFORMANCE CONCERNING INTERPRETIVE MORPHOLOGY FLAGGING OF THE LEUKOCYTE DIFFERENTIAL COUNT, Clinical and laboratory haematology, 19(3), 1997, pp. 187-190
Results from the Sysmex SF-3000 automated haematology analyser are rep
orted with special attention to leucocyte morphology Ragging in pathol
ogical conditions. Results for leucocyte differential counts provided
by the Sysmex SF-3000 were compared with those obtained from the Sysme
x NE-8000 and a visual evaluation based on a differential count of 400
leucocytes. One hundred samples from apparently healthy subjects and
100 samples from patients with haematological abnormalities were exami
ned, The SF-3000 yielded a rather high frequency of flagging inappropr
iately referring to blasts, immature granulocytes, left shift and nucl
eated red blood cells/platelet clumps, The presence of atypical lympho
cytes was not indicated by appropriate nagging.