EVALUATION OF SYSMEX SF-3000 PERFORMANCE CONCERNING INTERPRETIVE MORPHOLOGY FLAGGING OF THE LEUKOCYTE DIFFERENTIAL COUNT

Citation
Pcm. Bartels et al., EVALUATION OF SYSMEX SF-3000 PERFORMANCE CONCERNING INTERPRETIVE MORPHOLOGY FLAGGING OF THE LEUKOCYTE DIFFERENTIAL COUNT, Clinical and laboratory haematology, 19(3), 1997, pp. 187-190
Citations number
3
Categorie Soggetti
Hematology
ISSN journal
01419854
Volume
19
Issue
3
Year of publication
1997
Pages
187 - 190
Database
ISI
SICI code
0141-9854(1997)19:3<187:EOSSPC>2.0.ZU;2-7
Abstract
Results from the Sysmex SF-3000 automated haematology analyser are rep orted with special attention to leucocyte morphology Ragging in pathol ogical conditions. Results for leucocyte differential counts provided by the Sysmex SF-3000 were compared with those obtained from the Sysme x NE-8000 and a visual evaluation based on a differential count of 400 leucocytes. One hundred samples from apparently healthy subjects and 100 samples from patients with haematological abnormalities were exami ned, The SF-3000 yielded a rather high frequency of flagging inappropr iately referring to blasts, immature granulocytes, left shift and nucl eated red blood cells/platelet clumps, The presence of atypical lympho cytes was not indicated by appropriate nagging.