A. Nann et E. Pretsch, POTENTIOMETRIC DETECTION OF ANIONS SEPARATED BY CAPILLARY ELECTROPHORESIS USING AN ION-SELECTIVE MICROELECTRODE, Journal of chromatography, 676(2), 1994, pp. 437-442
Capillary electrophoresis in fused-silica capillaries of 10 mu m I.D.
coupled to an anion-selective microelectrode as on-column detector all
ows inorganic and organic anions, especially lipophilic ones, to be se
parated with high resolution and sensitivity of detection. For perchlo
rate, plate numbers of up to 10(7)/h and a detection limit of 5.10(-8)
M have been achieved. After delogarithmizing the response function of
the microelectrode, quantitative results are obtained as described pr
eviously.