LASER SAFETY ANALYSIS OF A RETINAL SCANNING DISPLAY SYSTEM

Citation
E. Viirre et al., LASER SAFETY ANALYSIS OF A RETINAL SCANNING DISPLAY SYSTEM, Journal of laser applications, 9(5), 1997, pp. 253-260
Citations number
11
Categorie Soggetti
Optics
ISSN journal
1042346X
Volume
9
Issue
5
Year of publication
1997
Pages
253 - 260
Database
ISI
SICI code
1042-346X(1997)9:5<253:LSAOAR>2.0.ZU;2-G
Abstract
The Virtual Retinal Display (VRD) is a visual display that scans modul ated laser light on to the retina of the viewer's eye to create an ima ge. Maximum permissible exposures (MPE) have been calculated for the V RD in both normal viewing and possible failure modes. The MPE power le vels are compared to the measured power that enters the eye while view ing images with the VRD. The power levels indicate that the VRD is saf e in both normal operating mode and in failure modes.