REPORT ON THE COOPERATIVE DETERMINATION OF MOLECULAR-WEIGHT AVERAGES OF POLYMERS BY SIZE-EXCLUSION CHROMATOGRAPHY .7. A REPORT ON THE FORTHROUND-ROBIN TEST (NO-1)
S. Mori et al., REPORT ON THE COOPERATIVE DETERMINATION OF MOLECULAR-WEIGHT AVERAGES OF POLYMERS BY SIZE-EXCLUSION CHROMATOGRAPHY .7. A REPORT ON THE FORTHROUND-ROBIN TEST (NO-1), Bunseki Kagaku, 46(10), 1997, pp. 837-844
Molecular weight averages of polystyrene (PS) samples (SRM 706, SRM 70
5a, and PS-1) were determined at 24 laboratories belonging to the sect
ion of special study on SEC, using two sets of PS calibration standard
s commercially available at Tosoh (TSK) and Showa Denko (Shodex) compa
nies. The PS-1 sample was used in the last three round robin tests. Th
e sample concentrations, injection volume and other experimental condi
tions were prescribed as at the 2nd round robin test. Two calibration
curves were constructed using TSK standards and Shodex standards, resp
ectively. Molecular weight averages of the three samples calculated us
ing the Shodex calibration curve were 3 to 14% higher than those calcu
lated using the TSK calibration curve, and the differences between two
molecular weight averages calculated using the Shodex and TSK calibra
tion curves were significant at the 1% significance level. This means
that when the values of the molecular weight averages are reported, th
e reporter has to declare which PS calibration standards were employed
.