VALIDATION OF THE TECHNICON H-ASTERISK-1 FOR THE AUTOMATED CELL COUNTAND CELL-DIFFERENTIATION OF FELINE BLOOD-SAMPLES

Citation
R. Liedl et J. Hirschberger, VALIDATION OF THE TECHNICON H-ASTERISK-1 FOR THE AUTOMATED CELL COUNTAND CELL-DIFFERENTIATION OF FELINE BLOOD-SAMPLES, Kleintierpraxis, 42(9), 1997, pp. 703
Citations number
8
Categorie Soggetti
Veterinary Sciences
Journal title
ISSN journal
00232076
Volume
42
Issue
9
Year of publication
1997
Database
ISI
SICI code
0023-2076(1997)42:9<703:VOTTHF>2.0.ZU;2-J
Abstract
The Technicon H1 gives a reliable CBC in the cat except the thrombocy te count. The differentiation of neutrophils and lymphocytes is accura te enough for clinical purposes. The differentiation of eosinophiles i s not accurate. The monocyte estimations of the Technicon H1 differ f rom those of the reference method, a 400 cell leucocyte differential c ell count. Storage of the blood at room temperature and at 4 degrees C is possible for three days, but storage at room temperature for seven days results in a significant decrease of neutrophils.