Mc. Rouan et al., AUTOMATED MICROANALYSIS OF OXCARBAZEPINE AND ITS MONOHYDROXY AND TRANSDIOL METABOLITES IN PLASMA BY LIQUID-CHROMATOGRAPHY, Journal of chromatography B. Biomedical applications, 658(1), 1994, pp. 167-172
Citations number
12
Categorie Soggetti
Chemistry Analytical
Journal title
Journal of chromatography B. Biomedical applications
An automated high-performance liquid chromatographic method for the si
multaneous determination of oxcarbazepine and its monohydroxy and tran
sdiol metabolites in plasma is described. ihydro-11-oxo-11H-dibenz[b,e
]azepine-5-carboxamide was used as internal standard. Liquid-solid ext
raction from plasma (100 mu l) on 50 mg Bond-Elut C-18 cartridges was
automatically performed by the Automatic Sample Preparation with Extra
ction Columns (ASPEC) system. A reversed-phase column (ODS Hypersil, 3
mu m particle size, 4 cm x 4.6 mm I.D.) was used with acetonitrile-me
thanol-0.01 M potassium dihydrogenphosphate as mobile phase. The elute
d compounds were detected at 210 nm. The limit of quantitation was 0.2
mu mol/l for oxcarbazepine and 0.1 mu mol/l for its metabolites. No i
nterference with concomitantly administered anti-epileptic drugs such
as phenobarbital, phenytoin, valproic acid or carbamazepine, was found
.