Ras. Cooper et al., INTERNAL CARDIOVERSION OF ATRIAL-FIBRILLATION - MARKED REDUCTION IN DEFIBRILLATION THRESHOLD WITH DUAL CURRENT PATHWAYS, Circulation, 96(8), 1997, pp. 2693-2700
Background The ultimate acceptance of a fully automatic atrial defibri
llator will depend on the reduction of pain to acceptable levels, requ
iring a marked decrease in defibrillation thresholds. The purpose of t
his study was to determine whether atrial defibrillation thresholds ca
n be reduced by sequential shocks delivered through two current pathwa
ys. Methods and Results Sustained atrial fibrillation was induced with
rapid atrial pacing in 12 adult sheep. Defibrillation electrodes were
positioned in the right atrial appendage (RAap), distal coronary sinu
s (DCS), proximal coronary sinus (CSos), main/left pulmonary artery ju
nction (PA), and right ventricular apex (RV). Single-capacitor biphasi
c waveforms (3/1 ms) were delivered through combinations of these elec
trodes. Probability-of-success curves were determined for single shock
s with a single current pathway and sequential shocks with either sing
le-or dual current pathways. The ED50 for delivered energy for the dua
l current pathway RAap to DCS then CSos to PA was 0.36+/-0.13 J, which
was significantly lower than the ED50 of the standard single current
pathway RAap to DCS (1.31+/-0.3 J) and was significantly lower than al
l other configurations tested. Conclusions Internal atrial defibrillat
ion thresholds can be markedly reduced with two sequential biphasic sh
ocks delivered over two current pathways compared with the standard si
ngle shock delivered over a single current pathway or with sequential
shocks delivered over a single current pathway.