Ma. Thornton et al., 3-DIMENSIONAL STRESS-ANALYSIS OF POLYPROPYLENE LEAFLETS FOR PROSTHETIC HEART-VALVES, Medical engineering & physics, 19(6), 1997, pp. 588-597
The effect of changing the modulus and thickness of the material in th
e leaflets of an artificial heart valve has been investigated. This ha
s been achieved with a finite element model of the valve having approx
imately 2300 thin shell elements. The valve motion and the resulting s
tresses are modelled dynamically during closure, and subsequent pressu
risation. The stresses decrease as the leaflets are made thicker and t
he modules is increased. Local and global thickening has been investig
ated. The highest stresses appear at the tops of the stent posts in th
e regions of the commissures. When modulus is too low, or the leaflets
are too thin, the valve prolapsed. (C) 1997 IPEM. Published by Elsevi
er Science Ltd.