PIEZOELECTRIC BIOSENSOR FOR DETECTION OF SALMONELLA-TYPHIMURIUM

Citation
Jm. Ye et al., PIEZOELECTRIC BIOSENSOR FOR DETECTION OF SALMONELLA-TYPHIMURIUM, Journal of food science, 62(5), 1997, pp. 1067
Citations number
17
Categorie Soggetti
Food Science & Tenology
Journal title
ISSN journal
00221147
Volume
62
Issue
5
Year of publication
1997
Database
ISI
SICI code
0022-1147(1997)62:5<1067:PBFDOS>2.0.ZU;2-1
Abstract
A flow injection analysis (FIA) system was developed based on a piezoe lectric biosensor for detection of Salmonella typhimurium. The anti-Sa lmonella spp antibody was immobilized onto the gold electrode coated q uartz crystal surface through a polyethylenimine-glutaraldehyde (PEG) technique and dithiobis-succinimidyl propionate (DSP) coupling. The PE G technique proved more successful for FIA applications than the DSP c oupling. The biosensor had responses of 23-47 Hz in 25 min when the PE G immobilization technique was employed, with R-2>0.94 for Salmonella typhimurium concentrations of 5.3x10(5) to 1.2x10(9) CFU/mL.