Jj. Curtis et al., IN-VIVO LEFT-VENTRICULAR ASSIST INDUCED COAGULATION DERANGEMENTS - COMPARISON OF SARNS-3M AND ST. JUDE MEDICAL CIRCUITS, ASAIO journal, 43(5), 1997, pp. 414-417
An in vitro comparison of centrifugal pumping systems manufactured by
Sarns-3M and St. Jude Medical revealed a difference in blood cell dera
ngement. The purpose of this study was to compare in vivo the effects
of 96 hr of left ventricular assist (LVA) on indexes of coagulopathy,
hemolysis, and complement activation. Two groups of calves (each: n =
5) were instrumented with identical left atrial to thoracic aorta cent
rifugal pumping circuits using either Sarns-3M or St. jude centrifugal
pumps. Laboratory evaluations were performed pre-assist and at 1, 4,
24, 48, 72, and 96 hr during LVA. Platelet counts dropped significantl
y by 24 hr (Sarns-3M: 28%; St. Jude: 30%); no significant change in fu
nction was noted. Activated clotting time increased slightly (p > 0.05
). Prothrombin time increased at 4 and 24 hr of LVA, returning to base
line by 96 hr (p < 0.05). Activated partial thromboplastin time increa
sed with the St. jude device from 24 to 96 hr on LVA (p < 0.05); the i
ncrease with the Sarns-3M device never reached significance. No signif
icant changes in lactate dehydrogenase or plasma free hemoglobin were
detected. Complement fraction C5a rose by 1 hr of LVA (p < 0.05), peak
ing at 4 hr and returning to baseline by 96 hr with both pumps. No sig
nificant difference was detected between pump groups for any of the pa
rameters. it was concluded that 1) 96 hr Sarns-3M and St. Jude LVA cau
sed coagulation derangement in calves, 2) neither pump demonstrated an
advantage regarding coagulation and complement parameters, 3) hemolys
is observed with the Sarns-3M pump in vitro was not evidenced in vivo,
and 4) in vitro evidenced centrifugal pump differences may not be rea
lized in vivo.