Zj. Cao et L. Qian, A TRIPLE-HEAD SPECT SYSTEM WITH PARALLEL-HOLE COLLIMATORS OF DIFFERENT ACCEPTANCE ANGLES, International journal of medical informatics, 46(3), 1997, pp. 145-157
Citations number
9
Categorie Soggetti
Information Science & Library Science","Medical Informatics
We proposed to use three different parallel-hole collimators for a tri
ple-head SPECT system. One of the collimators had a small collimator a
cceptance angle to provide ultra-high spatial resolution and the other
two had larger collimator acceptance angles to achieve high counts. A
new 2D reconstruction algorithm that combined the data acquired from
different collimators was derived to take advantages of both high reso
lution and high sensitivity. The algorithm was evaluated using a compu
ter-simulated matrix of spherical sources. For noise-free data, the ac
curacy (mainly determined by spatial resolution) obtained from combina
tion of the collimator acceptance angles of 1.35, 4.05 and 6.75 degree
s (or 1.35, 5.40 and 9.45 degrees) was slightly inferior to that obtai
ned from three same LEHR collimators (with a 2.70 degrees collimator a
cceptance angle). This is because the modulation transfer function (MT
F) resulting from three different collimators decreases more quickly a
t low frequencies but becomes comparable at high frequencies as compar
ed with the MTF of the 2.70 degrees collimator. With noisy data, howev
er, the image quality obtained with three different collimators was be
tter than that resulting from any combinations of three same collimato
rs. The improvement was only achieved by using the derived algorithm,
while the conventional FBP algorithm did not improve image quality eve
n with the same collimator configuration. (C) 1997 Elsevier Science Ir
eland Ltd.