COMPARISON OF GRAPHITE-TO-WATER ABSORBED-DOSE TRANSFERS FOR CO-60 PHOTON BEAMS USING IONOMETRY AND FRICKE DOSIMETRY

Citation
B. Chauvenet et al., COMPARISON OF GRAPHITE-TO-WATER ABSORBED-DOSE TRANSFERS FOR CO-60 PHOTON BEAMS USING IONOMETRY AND FRICKE DOSIMETRY, Physics in medicine and biology, 42(11), 1997, pp. 2053-2063
Citations number
19
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
00319155
Volume
42
Issue
11
Year of publication
1997
Pages
2053 - 2063
Database
ISI
SICI code
0031-9155(1997)42:11<2053:COGATF>2.0.ZU;2-1
Abstract
To derive the absorbed dose to water from a standard of absorbed dose to graphite, the metrology laboratories which apply such a method usua lly make use of cavity ionization chambers as transfer instruments. In addition, the BNM-LPRI has tested, as such instruments, two types of Fricke dosimeter in its cobalt-60 beam The two procedures are compared and their results are found to be in good agreement (the difference i s less than 0.1%). Both procedures are then taken into account for the calculation of the reference value of absorbed dose to water.