AC-227 - A SUGGESTED LONG-LIVED MULTIPLE-LINE GAMMA-RAY CALIBRATION STANDARD

Citation
J. Vanaarle et al., AC-227 - A SUGGESTED LONG-LIVED MULTIPLE-LINE GAMMA-RAY CALIBRATION STANDARD, Applied radiation and isotopes, 45(4), 1994, pp. 419-427
Citations number
9
Categorie Soggetti
Nuclear Sciences & Tecnology","Radiology,Nuclear Medicine & Medical Imaging
Journal title
Applied radiation and isotopes
ISSN journal
09698043 → ACNP
Volume
45
Issue
4
Year of publication
1994
Pages
419 - 427
Database
ISI
SICI code
0969-8043(1994)45:4<419:A-ASLM>2.0.ZU;2-P
Abstract
x-Ray and gamma-ray lines following the decay of Ac-227 in transient e quilibrium with its daughters were assayed via high-purity germanium-c rystal (HPGe) spectrometry of a chemically purified Ac-227 gamma-ray s ource. Relative emission probabilities were determined for 120 lines ( I(gamma) greater-than-or-equal-to 0.003%), and conversion to absolute emission probabilities was achieved via normalization to the known int ensity of the well-characterized 351-keV gamma-ray line of Bi-211, a d aughter of Ac-227. Special emphasis was paid to the stronger lines, wh ich can be used for energy and/or efficiency calibrations of germanium detectors.