This article presents case studies in which single-channel mapping was
used with patients who were experiencing discomfort while using their
cochlear implant devices. Repeated psycophysical testing together wit
h integrity testing had failed to locate the source of the problem in
each of the described cases. Single-channel mapping was then used as a
nother means of device troubleshooting. Single-channel maps were creat
ed for each electrode across the array. In each case, the patient was
able to identify the offending electrode(s) during the presentation of
speech stimuli, whereas the problem had not been evident during psych
ophysical testing with pulse stimuli. Eliminating these electrodes fro
m the map alleviated the problems experienced by these implant users i
n each case.