SINGLE ELECTRODE MAPS IN DEVICE TROUBLESHOOTING

Citation
Av. Hodges et al., SINGLE ELECTRODE MAPS IN DEVICE TROUBLESHOOTING, The American journal of otology, 18(6), 1997, pp. 115-115
Citations number
1
Categorie Soggetti
Otorhinolaryngology
ISSN journal
01929763
Volume
18
Issue
6
Year of publication
1997
Supplement
S
Pages
115 - 115
Database
ISI
SICI code
0192-9763(1997)18:6<115:SEMIDT>2.0.ZU;2-W
Abstract
This article presents case studies in which single-channel mapping was used with patients who were experiencing discomfort while using their cochlear implant devices. Repeated psycophysical testing together wit h integrity testing had failed to locate the source of the problem in each of the described cases. Single-channel mapping was then used as a nother means of device troubleshooting. Single-channel maps were creat ed for each electrode across the array. In each case, the patient was able to identify the offending electrode(s) during the presentation of speech stimuli, whereas the problem had not been evident during psych ophysical testing with pulse stimuli. Eliminating these electrodes fro m the map alleviated the problems experienced by these implant users i n each case.