The aim of this study is to review our experience with the Optifix por
ous-coated cementless acetabular component at an average 5 year follow
-up. In a retrospective clinical investigation, the amount and rate of
linear wear of the polyethylene of 26 cementless Optifix acetabular c
omponents implanted during a primary Hybrid (Spectron femoral componen
t) total hip replacement in 24 patients, were determined by a comparis
on of the thicknesses of the cup as measured on the initial and most r
ecent follow-up radiographs. The measurements were corrected for magni
fication error by direct measurement of the diameter of the femoral he
ad on each radiograph according to the technique of Livermore et al. T
he mean duration of follow-up was 60 months (range, 48 to 73 months).
Linear wear averaged 0.53 +/- 0.12 mm (0 to 2.28 mm) overall, with a m
ean rate of 0.11 +/- 0.02 mm (range 0 to 0.46 mm) per year p < 0.005.
No osteolysis of the acetabulum was seen. None of the acetabular compo
nents migrated. Osteolysis of the endosteum of the femur was found in
six hips. None of the femoral components subsided. In conclusion, wear
of the polyethylene in this cup has not been previously reported, and
is consistent with results published with other types of acetabular c
omponents.