Ch. Versteeg et al., IMPACT OF SCALE STANDARDIZATION ON IMAGES OF DIGITAL RADIOGRAPHY SYSTEMS, Dento-maxillo-facial radiology, 26(6), 1997, pp. 337-343
Objectives: To test the effect of altering image size on diagnostic qu
ality. Methods: Endodontic files, size 10 and 15 were prepared to full
root length and 1.5 mm short, in upper and lower molars and premolars
. Digital images of the Sidexis (S) and Visualix/Vixa (V) system were
reduced in size to Digora (D) scale, and the D images were enlarged to
S scale. ROC analysis was performed and data analysed with MANOVA sta
tistics. Results: Assessment of root canal length with file size 10 wa
s less accurate for the S images reduced in size, compared with the or
iginal S images (P < 0.011). All other differences were not significan
t (P > 0.081). Conclusions: Relevant diagnostic information may be los
t when images are reduced in size. Therefore, for optimal presentation
, the smaller images should be enlarged rather than the larger ones re
duced.