FLOW-INJECTION SPECTROPHOTOMETRIC DETERMINATION OF TOTAL CARBONYLS INHYDROCARBONS BY REACTION WITH VANILLIN

Citation
Jf. Vanstaden et He. Britz, FLOW-INJECTION SPECTROPHOTOMETRIC DETERMINATION OF TOTAL CARBONYLS INHYDROCARBONS BY REACTION WITH VANILLIN, Analytica chimica acta, 351(1-3), 1997, pp. 281-289
Citations number
12
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032670
Volume
351
Issue
1-3
Year of publication
1997
Pages
281 - 289
Database
ISI
SICI code
0003-2670(1997)351:1-3<281:FSDOTC>2.0.ZU;2-S
Abstract
The determination of total carbonyls in hydrocarbons by flow injection is based on the condensation reaction between the carbonyl group and vanillin in a strongly alkaline sodium hydroxide medium. The carbonyls , in an organic phase, were introduced into an aqueous carrier stream, mixed and extracted in a phase separator before the reaction product in the aqueous phase was detected spectrophotometrically at 414 nm. Pa rameters affecting the performance of the robust analyser were studied . The proposed analyser is fully computerised and is able to monitor t otal carbonyls in samples at a frequency of 15.3 samples per hour with a relative standard deviation <2.5% at ca 500 mg l(-1). The calibrati on graphs are linear between 100 and 1000 mg l(-1) for acetone and bet ween 250 and 1000 mg l(-1) for other carbonyl compounds. The detection limits for the individual carbonyl compounds are given. The proposed system is suitable for the determination of total carbonyl compounds i n organic solvents in the range 0.025-0.1% m/v calculated as acetone. (C) 1997 Elsevier Science B.V.