Jc. Barry et A. Backes, LIMBUS VERSUS PUPIL CENTER FOR OCULAR ALIGNMENT MEASUREMENT WITH CORNEAL REFLEXES, Investigative ophthalmology & visual science, 38(12), 1997, pp. 2597-2607
Purpose. To investigate the accuracy of ocular misalignment measuremen
t, using corneal reflections. Methods, Corneal reflex positions were m
easured relative to two landmarks, the limbus center and the entrance
pupil center, using high-resolution digital images for cyclopean gaze
angles from 0 degrees to 18.8 degrees (34.04 prism diopters; PD) to th
e right and to the left in 10 subjects. Distance h from the center of
the corneal curvature to each landmark was determined from linear regr
essions and showed significant differences between both conditions: me
an h(limbus) was 5.243 mm, and mean h(pupil) was 4.884 mm, From these,
limbus-and entrance-pupil-center-related Hirschberg ratios were deter
mined as 11 degrees/mm (19.43 PD/mm) and 11.82 degrees/mm (20.92 PD/mm
), respectively; and ocular alignment was calculated for both conditio
ns. Simulated angles of strabismus were calculated as the binocular di
fference between ocular alignment of the right and left eyes (conditio
n 1), and as the monocular difference between the ocular alignment of
right eyes and left eyes separately (condition 2). Results. Condition
1: Errors in simulated angles of strabismus were approximately twice a
s large for entrance pupil center compared with those in limbus-center
-based evaluation; in the primary position, the 95% pupil-related conf
idence interval of the binocular difference was +/-5.217 degrees (9.1
PD), compared with +/-3.174 degrees (5.5 PD) for the limbus-related op
tion. Condition 2: Errors were approximately equal. Conclusions. The e
ntrance pupil center is a less reliable landmark than is the limbus ce
nter for measuring ocular alignment by using corneal reflections, beca
use of unreliable positions of the entrance pupil center; different me
an Hirschberg ratios should be used in both conditions.