A simple procedure to analyze ammonia in ethylene by ion mobility spec
trometry is described. The spectrometer is operated with a silane poly
mer membrane, Ni-63 ion source, H+(H2O)(n) reactant ion, and nitrogen
drift and source gas. Ethylene containing parts per billion (ppb) (v/v
) concentrations of ammonia is pulled across the membrane and diffuses
into the spectrometer. Preconcentration or preseparation is unnecessa
ry, because the ethylene in the spectrometer has no noticeable effect
on the analytical results. Ethylene does not polymerize in the radioac
tive source. Ethylene's flammability is negated by the nitrogen inside
the spectrometer. Response to ammonia concentrations between 200 ppb
and 1.5 ppm is near linear, and a detection limit of 25 ppb is calcula
ted. (C) 1997 Elsevier Science B.V.