FREE-MASS RECIPROCITY CALIBRATION

Authors
Citation
Tb. Gabrielson, FREE-MASS RECIPROCITY CALIBRATION, The Journal of the Acoustical Society of America, 102(5), 1997, pp. 2800-2808
Citations number
12
Categorie Soggetti
Acoustics
ISSN journal
00014966
Volume
102
Issue
5
Year of publication
1997
Part
1
Pages
2800 - 2808
Database
ISI
SICI code
0001-4966(1997)102:5<2800:FRC>2.0.ZU;2-6
Abstract
A simple method for reciprocity calibration of accelerometers and velo city sensors is presented. The useful frequency range is from about 1 Hz to a kilohertz or more and the method is capable of subpercent accu racy. Calibrations with excitation levels from 1 to 300 mm/s(2) (0.1 m g to 0.03 g) were produced in prototype fixtures. The device to be cal ibrated and two geophones are used in the calibration fixture. Since t he geophones are both reciprocal, a direct test of reciprocity can be done prior to the calibration to evaluate the apparatus. In contrast t o standard techniques, the method requires only one relatively inexpen sive mechanical setup.