V. Gulani et al., A MULTIPLE ECHO PULSE SEQUENCE FOR DIFFUSION TENSOR IMAGING AND ITS APPLICATION IN EXCISED RAT SPINAL-CORDS, Magnetic resonance in medicine, 38(6), 1997, pp. 868-873
A new imaging sequence for rapid determination of the apparent self-di
ffusion tensor of water was developed and tested on fixed excised rat
spinal cords. To reduce the time required to determine the tensor, the
sequence utilized a new single-shot approach with multiple spin echoe
s. An assumption of cylindrical symmetry in the sample was made, thus
requiring the measurement of only four of the six unique elements of t
he tenser. This assumption was found experimentally to be valid, and t
he results obtained using tile new sequence were found to be quantitat
ively the same as results obtained using a standard spin-echo sequence
.