ELIMINATION OF EDDY-CURRENT ARTIFACTS IN DIFFUSION-WEIGHTED ECHO-PLANAR IMAGES - THE USE OF BIPOLAR GRADIENTS

Citation
Al. Alexander et al., ELIMINATION OF EDDY-CURRENT ARTIFACTS IN DIFFUSION-WEIGHTED ECHO-PLANAR IMAGES - THE USE OF BIPOLAR GRADIENTS, Magnetic resonance in medicine, 38(6), 1997, pp. 1016-1021
Citations number
23
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
07403194
Volume
38
Issue
6
Year of publication
1997
Pages
1016 - 1021
Database
ISI
SICI code
0740-3194(1997)38:6<1016:EOEAID>2.0.ZU;2-N
Abstract
Small gradient fields resulting from incompletely canceled eddy curren ts can cause geometric distortion in echo-planar images. Although this distortion is negligible in most echo-planar applications, the large gradient pulses used in diffusion-weighted echo-planar imaging can res ult in significant image distortion. In this report, it is shown that this distortion can be significantly reduced by the application of bip olar gradient waveforms. Both bipolar diffusion-sensitizing gradients and an inverted gradient preparatory pulse were examined for minimizin g the eddy currents responsible for these distortions.