H. Chao et al., MULTI-ECHO P-31 SPECTROSCOPIC IMAGING OF ATP - A SCAN TIME REDUCTION STRATEGY, Journal of magnetic resonance imaging, 7(2), 1997, pp. 425-433
Spectroscopic imaging of P-31 metabolites and adenosine triphosphate (
ATP) in particular with multiple spin echoes may prove useful for redu
cing data acquisition times, The usual T2 decay processes that degrade
multi-echo spectroscopic imaging methods, however, are further compou
nded by J-coupling modulations in the case of ATP, We determine how th
ese modulations affect multi-echo spectroscopic imaging k-space data a
nd produce systematic spatial misregistrations of the ATP resonances,
The specific J-coupling modulations of ATP are determined to identify
echo-spacing effects in multi-echo spectroscopic imaging of ATP and to
determine appropriate post-processing correction schemes to address t
he spatial misregistration problem, An in vivo demonstration of the te
chnique that offers a threefold reduction in scan time compared to con
ventional SI methods is provided and compared with the conventional SI
approach.