Cyl. Hsu et Rr. Walters, ASSAY OF BROPIRIMINE IN RAT PLASMA BY MEANS OF ROBOTIC SOLID-PHASE EXTRACTION AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY, Journal of chromatography, 762(1-2), 1997, pp. 243-249
Citations number
4
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
A rapid and simple high-performance liquid chromatography (HPLC) metho
d incorporating automated solid-phase extraction (SPE) is described fo
r the determination of bropirimine, 2-amino-5-bromo-6-phenyl-4(3H)-pyr
imidinone, in plasma samples from various species. Using an automated
sample processor, plasma samples were loaded onto C-18 SPE columns and
the drug eluted with ethanol-methylene chloride (10:90, v/v). The ext
racts were analyzed by reversed-phase HPLC using a C, column with a mo
bile phase consisting of acetonitrile-water-trifluoroacetic acid (20:8
0:0.1, v/v/v). The UV absorbance of the column effluent was monitored
at a wavelength of 292 nm. Linear calibration curves were obtained in
the range of 0.01 to 22 mu g/ml. Precision (less than or equal to 4% r
elative standard deviation) and accuracy (less than or equal to 5% err
or) were acceptable at the concentrations evaluated. Application of th
is method to the quantitation of bropirimine in rat plasma for a toxic
okinetic/bioavailability study is reported.