A HIGH-PRECISION MANOMETRIC SYSTEM FOR ABSOLUTE CALIBRATIONS OF CO2 IN DRY AIR

Citation
Cl. Zhao et al., A HIGH-PRECISION MANOMETRIC SYSTEM FOR ABSOLUTE CALIBRATIONS OF CO2 IN DRY AIR, JOURNAL OF GEOPHYSICAL RESEARCH-ATMOSPHERES, 102(D5), 1997, pp. 5885-5894
Citations number
14
Categorie Soggetti
Metereology & Atmospheric Sciences
Volume
102
Issue
D5
Year of publication
1997
Pages
5885 - 5894
Database
ISI
SICI code
Abstract
A high-precision manometric system has been developed for absolute cal ibrations of CO2-in-air mixture gas. This report describes the princip le of the calibration method and evaluates the performance of the mano metric apparatus. The test results show that the reproducibility of th e manometric system for calibrating CO2-in-air gas mixtures is within +/-0.1 mu mol mol(-1) in the atmospheric CO2 concentration range of 30 0 to 400 mu mol mol(-1). Preliminary measurements indicate that the ag reement with the absolute World Meteorological Organization mole fract ion scale, widely used for atmospheric CO2 monitoring, is also within +/-0.1 mu mol mol(-1) in that same range.