SEVERE HEMOLYSIS DUE TO ARTIFICIAL CHORDAE DISPLACEMENT

Citation
S. Kihara et al., SEVERE HEMOLYSIS DUE TO ARTIFICIAL CHORDAE DISPLACEMENT, Journal of heart valve disease, 6(1), 1997, pp. 69-70
Citations number
6
Categorie Soggetti
Cardiac & Cardiovascular System
ISSN journal
09668519
Volume
6
Issue
1
Year of publication
1997
Pages
69 - 70
Database
ISI
SICI code
0966-8519(1997)6:1<69:SHDTAC>2.0.ZU;2-8
Abstract
Hemolytic anemia is a well-recognized complication of mechanical heart valve prosthesis but, as yet, has not been reported after mitral valv e repair with chordal replacement. We report a case of severe hemolyti c anemia after mitral valve repair with chordal replacement and Carpen tier-Edwards annuloplasty ring insertion. Progressive prolapse of the anterior leaflet due to the artificial chordae being too long caused r ecurrent regurgitation which was responsible for the hemolysis. The pa tient also had idiopathic thrombocytopenic purpura, but successful sec ond mitral valve repair was performed after high-dose gamma-globulin t herapy.