Sw. Wise et al., TECHNICAL FACTORS OF CT ANGIOGRAPHY STUDIED WITH A CAROTID-ARTERY PHANTOM, American journal of neuroradiology, 18(3), 1997, pp. 401-408
Citations number
11
Categorie Soggetti
Clinical Neurology","Radiology,Nuclear Medicine & Medical Imaging
PURPOSE: To evaluate scanning parameters (conventional versus spiral C
T, section thickness, and pitch) and vessel orientation in the perform
ance of CT angiography, METHODS: Conventional CT and 1.0-, 1.5-, and 2
.0-pitch spiral CT acquisitions of a carotid phantom designed with ves
sels oriented parallel to the z-axis, 45 degrees oblique, and perpendi
cular to the z-axis were obtained with section thicknesses of 2, 4, an
d 8 mm. The phantom contained 32 vessels with 0% to 100% stenoses, Nor
mal and stenotic luminal diameters were measured and the number of art
ifacts was assessed, RESULTS: No overall difference was observed among
conventional and spiral CT acquisitions obtained with pitches of 1.0,
1.5, and 2.0. With thicker sections, CT angiographic accuracy decreas
ed and artifacts increased. The three-vessel orientations were relativ
ely comparable in accuracy in terms of the percentage of stenosis meas
ured. Vessels parallel to the z-axis suffered less artifactual degrada
tion. Unique artifacts, such as luminal distortion and beam hardening,
were observed in vessels oriented at 45 degrees and perpendicular to
the z-axis. CONCLUSION: Use of thinner sections with vessels oriented
parallel to the z-axis optimizes CT angiographic quality. There is no
apparent degradation with the use of spiral CT, and a pitch of 1.5 or
2.0 provides results equivalent to 1.0-pitch spiral studies.