H. Schwerdt et al., METHODOLOGICAL AND CLINICAL ASPECTS OF MI NIMIZING POWER-CONSUMPTION IN PACEMAKERS PROVIDING TELEMETRY, Herz, Kreislauf, 26(4), 1994, pp. 119-124
In order to minimize power consumption, 30 leads implanted in 17 patie
nts with a DDD-pacemaker were investigated using pacemaker telemetry.
At charge threshold and at output settings doubling charge threshold,
telemetric measurements of pulse amplitude, pulse duration, charge per
pulse and average cell current were carried out for voltages ranging
from 0.5-4 V. From the mean values of these data, we calculated the rh
eobase and chronaxie of the strength-duration curve and determined min
imum charge threshold. At a safety margin of 100% of threshold charge,
pulse amplitudes with minimum battery drain (U(opt) were linearly rel
ated to minimum voltages assuring capture (r = 0.78) and to the rheoba
se (r = 0.82). The relative frequency distribution of U(opt) showed pe
aks at singular pulse amplitudes. Therefore, programming a safety marg
in for pacing by doubling threshold charge should use the lowest pulse
amplitude feasible, taking into account advantageous voltages due to
output circuitry. Using Lapicque's law to double voltage above thresho
ld almost doubles pulse durations (factor 1.8 +/- 0.8) compared to the
charge method, increases battery current drain and is therefore not r
ecommended.