ELEMENTAL COMPOSITION DETERMINATION OF ORGANOPHOSPHORUS COMPOUNDS USING GAS-CHROMATOGRAPHY AND ATOMIC-EMISSION SPECTROMETRIC DETECTION

Citation
G. Becker et al., ELEMENTAL COMPOSITION DETERMINATION OF ORGANOPHOSPHORUS COMPOUNDS USING GAS-CHROMATOGRAPHY AND ATOMIC-EMISSION SPECTROMETRIC DETECTION, Analytica chimica acta, 340(1-3), 1997, pp. 181-189
Citations number
25
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032670
Volume
340
Issue
1-3
Year of publication
1997
Pages
181 - 189
Database
ISI
SICI code
0003-2670(1997)340:1-3<181:ECDOOC>2.0.ZU;2-D
Abstract
The elemental responses for a series of alkylated and arylated phospha tes evaluated were measured on an atomic emission spectrometric detect or. The signals for carbon, hydrogen, chlorine, phosphorus and oxygen were used to determine the specifity of elemental response ratios to a particular compound structure and the dependence of response ratios o n the amounts of the analytes. Variations of the response ratios, as w ell as the accuracy and variations of the stoichiometric values of the calculated empirical formulae, were evaluated. For this purpose, a te st mixture with reference compounds was used. Determination of empiric al formulae with acceptable errors of a few percent is possible if the calibrating reference substance is closely related - by structure, el emental composition, molecular weight and amount - to the compounds to be identified. Analyte amounts of at least 30 ng are required for eff icient calibration.