RESPONSE OF ALANINE DOSIMETERS AT VERY HIGH-DOSE-RATE

Citation
H. Kudoh et al., RESPONSE OF ALANINE DOSIMETERS AT VERY HIGH-DOSE-RATE, Applied radiation and isotopes, 48(4), 1997, pp. 497-499
Citations number
21
Categorie Soggetti
Nuclear Sciences & Tecnology","Radiology,Nuclear Medicine & Medical Imaging","Chemistry Inorganic & Nuclear
Journal title
Applied radiation and isotopes
ISSN journal
09698043 → ACNP
Volume
48
Issue
4
Year of publication
1997
Pages
497 - 499
Database
ISI
SICI code
0969-8043(1997)48:4<497:ROADAV>2.0.ZU;2-2
Abstract
Alanine dosimeters were irradiated at 3 x 10(10) Gy/s using a pulsed e lectron beam. The response measured by ESR showed good agreement with dosimetry by radiochromic film dosimeters, which are reliable at that absorbed dose rate. The results indicate that alanine dosimeters may b e used at very high dose rate, within their nominal-usable dose range up to 100 kGy. (C) 1997 Published by Elsevier Science Ltd.