RETARDING-FIELD ANALYZER USED IN ELASTIC PEAK ELECTRON-SPECTROSCOPY

Citation
D. Zeze et al., RETARDING-FIELD ANALYZER USED IN ELASTIC PEAK ELECTRON-SPECTROSCOPY, Vacuum, 48(3-4), 1997, pp. 399-401
Citations number
14
Categorie Soggetti
Physics, Applied
Journal title
VacuumACNP
ISSN journal
0042207X
Volume
48
Issue
3-4
Year of publication
1997
Pages
399 - 401
Database
ISI
SICI code
0042-207X(1997)48:3-4<399:RAUIEP>2.0.ZU;2-E
Abstract
The percentage, eta(e), of electrons elastically reflected by a substr ate is a fundamental parameter for elastic peak electron spectroscopy (EPES). A Retarding Field Analyser with ifs large acceptance angle can be used for this spectroscopy. However, distortion is observed in the low energy side of the recorded elastic peak, increasing with the ene rgy of the impinging electrons, which introduces difficulty for eta(e) measurement. To eliminate this difficulty, we propose in this paper a method to obtain eta(e) using a reference curve. The process is appli ed to the study of the Ag/Mo system, as well as experimental determina tion of the mean free path of electrons in indium samples. (C) 1997 El sevier Science Ltd.