Cm. Putman et al., TECHNICAL FEASIBILITY AND PERFORMANCE STUDIES OF A 0.009-INCH NITINOLMICROGUIDEWIRE FOR POTENTIAL NEUROENDOVASCULAR APPLICATIONS, Investigative radiology, 32(4), 1997, pp. 241-247
RATIONALE AND OBJECTIVES. The authors conduct technical feasibility an
d performance studies on a new 0.009-inch nitinol microguidewire for p
otential neuroendovascular applications. METHODS. In vivo microcathete
rizations of brachiocephalic arteries were performed in four pigs usin
g the 0.009-inch nitinol microguidewire with a commonly used microcath
eter and compared with the performance of commercially available micro
guidewires. Vessels catheterized using the 0.009-inch wire were evalua
ted by histopathologil analysis. A standardized, bench-top method of e
valuating basic mechanical properties of microguidewires also was used
to compare the 0.009-inch nitinol microguidewire with two commonly us
ed microguidewires. RESULTS. The 0.009-inch microguidewire had similar
steerability, tractability, torque control, and distal tip flexibilit
y to the commonly used microguidewires in the in vivo simulations, Ben
ch top testing showed the 0.009-inch microguidewire to have comparable
distal tip flexibility and objective mechanical properties to the com
monly used microguidewires. CONCLUSIONS. The comparable subjective and
objective mechanical properties of the 0.009-inch nitinol microguidew
ire to that of commonly used microguidewires further establishes the p
ossibility of clinical implementation.