DIRECT ANALYSIS OF CORTISOL AND CORTISONE IN HUMAN URINE BY SEMI-MICROCOLUMN LIQUID-CHROMATOGRAPHY WITH MIXED-FUNCTION PRECOLUMN
Citation
T. Kanda et al., DIRECT ANALYSIS OF CORTISOL AND CORTISONE IN HUMAN URINE BY SEMI-MICROCOLUMN LIQUID-CHROMATOGRAPHY WITH MIXED-FUNCTION PRECOLUMN, The Journal of microcolumn separations, 7(5), 1995, pp. 445-449
Categorie Soggetti
Chemistry Analytical
SICI code
1040-7685(1995)7:5<445:DAOCAC>2.0.ZU;2-0
Abstract
A column-switching system based on semi-microcolumns is constructed fo
r direct analysis of cortisol and cortisone in human urine samples. Ur
ine samples (100 mu l) were directly injected into a polymer-coated mi
xed-function (PCMF) column (35 mm x 1.5 mm i.d.), which removes protei
ns and fractionates an analyte-containing zone. The analyte-containing
fraction was thereafter transferred to a C18 semi-microcolumn (250 mm
x 1.5 mm i.d.). The analytes were monitored with ultraviolet (UV) abs
orption at 242 nm, with the detection limit for cortisol and cortisone
, 3 pmol/mL (at a S/N ratio of 5) for each. The system showed quantita
tive results for dilute standard samples using human urine as a solven
t. The method could be applied to the routine assessment of cortisol a
nd cortisone in human urine. (C) 1995 John Wiley & Sons, Inc.