TESTS OF CROSSED-WIRE POSITION-SENSITIVE PHOTOMULTIPLIERS FOR SCINTILLATING-FIBER PARTICLE TRACKING

Citation
Cf. Perdrisat et al., TESTS OF CROSSED-WIRE POSITION-SENSITIVE PHOTOMULTIPLIERS FOR SCINTILLATING-FIBER PARTICLE TRACKING, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 365(2-3), 1995, pp. 378-384
Citations number
9
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
365
Issue
2-3
Year of publication
1995
Pages
378 - 384
Database
ISI
SICI code
0168-9002(1995)365:2-3<378:TOCPPF>2.0.ZU;2-X
Abstract
Several applications of two Hamamatsu position sensitive photomultipli er tubes to the detection of high energy particles with scintillating fibers are discussed. The PMTs are of the multiwire anode grid design, type R2486 and R4135. These tubes were tested with both single sample s and arrays of 2 and 3 mm diameter scintillating fibers. Measurements of position resolution of the PMTs using either the charge digitizati on or the delay line readout techniques were made. The results indicat e an intrinsic inability of the technique to reconstruct the actual po sition of a fiber on the photocathode when its location falls halfway between two grid wires, A way to overcome this limit is suggested.