Sq. Xu et al., RELATIONSHIP BETWEEN CORONARY ANGIOSCOPIC AND INTRAVASCULAR ULTRASOUND IMAGING AND RESTENOSIS, Chinese medical journal, 108(10), 1995, pp. 743-749
In order to investigate the relationship between restenosis and the mo
rphology detected by coronary angioscopy (CASC) and introvascular ultr
asound imaging (IVUS), 17 patients were detected by CASC and IVUS imme
diately and 3 months after percutaneous transluminal coronary angiosco
py (PTCA), The results showed that the dilation index by IVUS (DIu) wa
s significantly lower in restenosis patients than in non-restenosis pa
tients (0.42 +/- 0.08 versus 0.78 +/- 0.16, P < 0.01) and that the ela
stic recoil (ER) was higher in restenosis patients than in non-resteno
sis patients (4.51 +/- 1.42 mm(2) versus 1.63 +/- 1.20 mm(2), P < 0.01
), and that the elastic recoil rate (ERR) was also higher in restenosi
s patients than in non-restenosis patients (57.3 +/- 8.07% versus 21.8
0 +/- 16.84% P < 0.01), and that coronary dissection, atheromatous pla
que and calcification as well as the colour of inner coronary artery h
ad no relation with chronic restenosis. In conclusion, the elastic rec
oil is one of the important factors of chronic restenosis after PTCA.