Purpose. Present videokeratographs provide corneal radius of curvature
measurements in terms of axial (sagittal) radius and, for some instru
ments, instantaneous (tangential) radius. The EyeSys videokeratograph
allows conversion from axial to instantaneous radius based on an algor
ithm, which assumes a basic corneal shape of elliptical form. Method.
We tested the accuracy of the EyeSys instantaneous radius algorithm fo
r 29 keratoconic corneas using the criterion that instantaneous radius
of the cone apex could be determined accurately by aligning the cone
apex with the optic axis of the videokeratoscope. Results. There was a
close relation between the instantaneous values obtained from the reg
ular and apex alignments for the 17 eyes with cone apices below 55 D (
95% limits of agreement -0.39 to 1.49 D), but not for the 12 eyes abov
e 55 D (95% limits of agreement -4.87 to 6.61 D). The mean distance of
the cone position from the center of the map was 1.16 mm for the inst
antaneous map and 1.59 mm for the axial map. Conclusion. Assuming the
criterion for accuracy to be the corneal radii found when the videoker
atograph was aligned with the cone apex, the instantaneous map for the
EyeSys videokeratograph at regular alignment provided reasonably vali
d measurements for corneas up to 55 D, but large errors existed for so
me corneas of higher power.