Since leakage of an expander dome is a well-known complication that ma
y lead to interruption of a tissue expansion procedure, the aim of thi
s project was to study different expansion domes in relation to a medi
cation port after perforation with various needles. Material and metho
ds: The combinations of five different expansion domes and one medicat
ion port were tested with three different needle types having two diff
erent diameters. The force-distance curves during puncturing, the punc
ture forces required and the leakage opening pressures were measured a
fter single on repeated punctures with small and large forces. A total
of more than 2000 experiments were performed. Results: Only one of th
e expander domes and the medication port had sufficient pressure stabi
lity. The technical characteristics of these devices are presented. Ho
wever, the different needle types and puncture techniques had only a m
inor influence on the pressure stability of the expander domes.