IMPROVED RESPONSE MATRICES OF BONNER SPHERE SPECTROMETERS WITH (LII)-LI-6 SCINTILLATION DETECTOR AND HE-3 PROPORTIONAL COUNTER BETWEEN 15 AND 100 MEV

Citation
V. Mares et H. Schraube, IMPROVED RESPONSE MATRICES OF BONNER SPHERE SPECTROMETERS WITH (LII)-LI-6 SCINTILLATION DETECTOR AND HE-3 PROPORTIONAL COUNTER BETWEEN 15 AND 100 MEV, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 366(1), 1995, pp. 203-206
Citations number
10
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
366
Issue
1
Year of publication
1995
Pages
203 - 206
Database
ISI
SICI code
0168-9002(1995)366:1<203:IRMOBS>2.0.ZU;2-U
Abstract
The neutron responses of Bonner sphere spectrometric systems with a 4 mm X 4 mm (LiI)-Li-6 scintillator and with a He-3 (32 mm diameter) pro portional counter were calculated using the MCNP Monte Carlo code for the energy range 14.8 to 100 MeV. Calculated responses have been inter polated to evaluate the response matrix for both systems; data tables are available on diskette.