J. Macek et P. Ptacek, DETERMINATION OF OFLOXACIN IN HUMAN PLASMA USING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND FLUORESCENCE DETECTION, Journal of chromatography B. Biomedical applications, 673(2), 1995, pp. 316-319
Citations number
8
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Journal title
Journal of chromatography B. Biomedical applications
A new method for the determination of ofloxacin in human plasma was de
veloped. Plasma proteins were precipitated with acetonitrile, the supe
rnatant concentrated and injected into a reversed-phase C-18 column. E
noxacin was used as an internal standard. The fluorimetric detection w
as performed at 282 nm for excitation and 450 nm for emission. Limit o
f quantitation was 20 ng/ml and the calibration curve was linear up to
6900 ng/ml.