HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF THE SILICON PHTHALOCYANINE PC-4 IN HUMAN BLOOD

Citation
Mm. Zuk et al., HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF THE SILICON PHTHALOCYANINE PC-4 IN HUMAN BLOOD, Journal of chromatography B. Biomedical applications, 673(2), 1995, pp. 320-324
Citations number
15
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Journal title
Journal of chromatography B. Biomedical applications
ISSN journal
15726495 → ACNP
Volume
673
Issue
2
Year of publication
1995
Pages
320 - 324
Database
ISI
SICI code
Abstract
A sensitive and reproducible method has been developed for the measure ment of the silicon phthalocyanine Pc4 in red blood cell concentrates (RBCC). The procedure involves extraction of the RBCC with acetonitril e, purification of the extracts with reversed-phase Sep-Pak C-18 cartr idges and determination of Pc 4 in the extracts by high-performance li quid chromatography (HPLC) using a reversed-phase C-18 column. The det ection limit with I-mi RBCC samples is 2 ng. This method is applicable to monitoring Pc 4 during its use as a photosensitizer for the inacti vation of viruses in RBCC prior to transfusion. It has the potential t o be adapted for measuring Pc 4 in tissues during its use in photodyna mic therapy of cancer.