WORKLOAD AND USE FACTOR OF MEDICAL LINEAR ACCELERATORS IN RADIOTHERAPY

Citation
T. Kron et al., WORKLOAD AND USE FACTOR OF MEDICAL LINEAR ACCELERATORS IN RADIOTHERAPY, Health physics, 69(6), 1995, pp. 971-975
Citations number
3
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
00179078
Volume
69
Issue
6
Year of publication
1995
Pages
971 - 975
Database
ISI
SICI code
0017-9078(1995)69:6<971:WAUFOM>2.0.ZU;2-E
Abstract
An important factor in the design of primary protective barriers is th e use factor, The present study was aimed at obtaining historical data on the use factor of two dual modality linear accelerators in a radio therapy department, Gantry angle, field size, and beam modifiers were recorded for all radiation qualities in use at two medical linear acce lerators with 6 MV and 18 MV x-rays and multiple electron energies ran ging from 4 MeV to 20 MeV, The data for one year of clinical use was e xtracted from a record and verifying system and an estimate of the phy sics workload on the machines was obtained by going through the qualit y assurance records and machine log books, Of the total dose of approx imately 37,000 Gy delivered in one year at isocenter on each unit 80% was given as 6 MV x-rays, As can be expected, most x-ray beams were di rected at the four cardinal gantry angles with the angular distributio n for 6 MV and 18 MV x-rays being very similar, Electron fields were b roadly distributed around the gantry pointing down position, Less than 25% of all clinical x-rag treatment fields extended beyond a field si ze of 200 cm(2).