Hi. Bak et al., PEAK ENERGY SHIFT DEPENDENT ON SOURCE POSITION IN GAMMA-RAY ENERGY MEASUREMENT BY USING A CLOSED-ENDED COAXIAL HPGE DETECTOR, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 366(2-3), 1995, pp. 332-339
The source position effect in the gamma-ray energy measurement by usin
g a closed-ended coaxial HPGe detector is investigated at the energy o
f 1332 keV. The shift of the gamma-ray peak energy is measured as a fu
nction of source-to-detector distance for axial and radial positions o
f the Co-60 source. By using Monte Carlo method, the interaction of ga
mma-rays in the detector material is simulated, and then the peak shif
t due to field increment effect and charge trapping effect is calculat
ed. The measured shift reasonably agrees with the calculated value. Fo
r the axial direction, the positive peak shift occurs, and the shift i
ncreases significantly within 3 cm from the detector. At the distance
of 0.5 cm, the measured shift becomes 42 eV. For the radial direction,
the negative peak shift within 9 eV occurs. The dependence of peak sh
ift on the source-to-detector distance for the radial direction is muc
h weaker than that for the axial direction.