E. Pocurull et al., IMPROVEMENT OF ONLINE SOLID-PHASE EXTRACTION FOR DETERMINING PHENOLIC-COMPOUNDS IN WATER, Chromatographia, 41(9-10), 1995, pp. 521-526
Citations number
21
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Modifying the most common design for the on-line coupling of a precolu
mn to reversed phase LC with diode array detection has resulted in red
uction of the broadening of the peaks which results when the compounds
of interest are strongly retained by a highly hydrophobic sorbent. Th
e modification consists of the desorption of the analytes trapped on t
he precolumn solely by the organic solvent used to modify the solvent
strength of the mobile phase. Results obtained using this design were
compared with those obtained with the conventional design, with C-18 a
nd PLRP-S precolumns. The performance of the system was also tested wi
th a highly cross-linked styrene-divinylbenzene copolymer (ENVI-chrom
P) precolumn for the determination of phenolic compounds in real sampl
es. The advantages and disadvantages are discussed. Ion-pair solid pha
se extraction is used in order to increase the breakthrough volumes of
more polar compounds, mainly phenol. The use of the new design enable
s phenolic compounds to be determined at the low mu g L(-1) level with
limits of detection ranging between 0.1 and 2 mu g L(-1) in tap water
when a 10 mL sample was analyzed.