DEVICE FOR EFFICIENT SOLVENT COLLECTION OF ENVIRONMENTALLY RELEVANT COMPOUNDS IN OFF-LINE SFE

Citation
Bw. Wenclawiak et al., DEVICE FOR EFFICIENT SOLVENT COLLECTION OF ENVIRONMENTALLY RELEVANT COMPOUNDS IN OFF-LINE SFE, Analytical chemistry, 67(24), 1995, pp. 4577-4580
Citations number
22
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032700
Volume
67
Issue
24
Year of publication
1995
Pages
4577 - 4580
Database
ISI
SICI code
0003-2700(1995)67:24<4577:DFESCO>2.0.ZU;2-Z
Abstract
The use of a Dewar condenser placed on top of the supercritical fluid extraction off-line collection vial reduces analyte losses signifcantl y. Five groups of analytes (EPA-PAHs, alkanes, nitro-PAHs, fatty acid esters, and chlorinated hydrocarbons), all of them of environmental co ncern, were investigated and quantified by GC-FID/GC-ECD. Collection l osses were reduced to about 2%, with the condenser kept at -25 degrees C.