A. Frazier et al., EFFECTS OF TREATMENT SETUP VARIATION ON BEAMS EYE VIEW DOSIMETRY FOR RADIATION-THERAPY USING THE MULTILEAF COLLIMATOR VS THE CERROBEND BLOCK, International journal of radiation oncology, biology, physics, 33(5), 1995, pp. 1247-1256
Citations number
13
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
Purpose: The purpose of this study is to quantify and compare retrospe
ctively the effects of treatment setup variation on beam's eye view (B
EV) dosimetry for radiation therapy using a multileaf collimator (MLC)
vs. cerrobend block. Methods and Materials: A study was performed on
a group of 18 patients with cancer of the head and neck, lung, and pel
vis who were treated with irregularly shaped fields. The BEV dosimetry
of the fields shaped with cerrobend blocks and the MLC was measured w
ith films at the depth of dose prescription in a solid water phantom.
A ''one-half-leaf'' insertion convention was used to shape the MLC. In
addition, an average of 15 sequential daily port films was taken per
patient during the course of radiotherapy. The port films were aligned
with the prescription film for each patient, Systematic error and ran
dom error of treatment setup for each patient were calculated. The eff
ects of setup variation were incorporated by convolving the patient po
rtal imaging data with the corresponding BEV film dosimetry. Two param
eters were used to quantify the BEV dosimetry. First, the field penumb
ra width was calculated, which represented the average of the normal s
eparations between 20 and 80% isodose lines along the prescription out
line. Second, the ratio of areas covered by the 90 and 20% isodose Lin
es, A90/20, was determined. The BEV dosimetry was then characterized w
ith and without the effects of treatment setup variation. In addition,
the difference in BEV dosimetry between the cerrobend block and the M
LC was used to estimate the corresponding changes in tumor control pro
bability (TCP). These changes were also compared to the changes in TCP
for the treatment with or without the effects of random setup variati
on. Results: With or without daily setup variation, the use of cerrobe
nd block was more favorable than the MLC in terms of the field penumbr
a width and A90/20 for all treatment sites. In the absence of daily va
riation, the MLC field penumbra width was on average 1.3 mm larger tha
n that of the cerrobend block, and 0.9 mm larger in the presence of da
ily setup variation. Similarly, the ratio A90/20 of the cerrobend bloc
k was on average 0.03 larger than that of MLC without daily setup vari
ation, and 0.02 with daily setup variation. The difference in field pe
numbra width and A90/20 between the MLC and the cerrobend block was sl
ightly reduced due to the effects of daily setup variation. For both t
he cerrobend block and the MLC, daily setup variation produced a signi
ficant increase in the held penumbra width, 2.3 mn for the cerrobend b
lock and 1.9 mm for the MLC, and a decrease in the A90/20, 0.06 for th
e former and 0.05 for the latter. The change due to the daily setup va
riation was about a factor of 2 larger than the changes due to replaci
ng the cerrobend block with the MLC. Using the TCP model, the change i
n TCP due to the daily setup variation was more than a factor of 3 lar
ger than the change in TCP due to replacing the cerrobend block with t
he MLC. It was noted that the average changes in the penumbra, the A90
/20 and the TCP calculated for the patient population did not adequate
ly describe the changes for the individual patient. Conclusion: Our re
sults do not show significant dosimetric differences between the MLC a
nd the cerrobend block in conventional radiation treatment, whether or
not daily setup variation was taken into consideration. The effects o
f daily setup variation alone produced a larger dosimetric change. The
same results were obtained when the data were applied to calculate ch
anges in TCP. For optimal radiation therapy, efforts should be concent
rated on reducing daily setup variation. Our results also demonstrate
the importance of frequent evaluation of MLC treatment using electroni
c portal imaging devices.